30
8Maintenanceandfaultrectication
PLICSMOBILE T61 • External GSM/GPRS radio unit
37700-EN-130228
Error code Cause Rectication
E042 Hardware error with
selfcheck
Send instrument for repair
E053 Sensor measuring
range not read cor-
rectly
Communication error: Check sensor
cable and screening
E086 Error communication
hardware (initiali-
sation of the radio
module failed)
Initialisation is carried out automati-
cally. If the error exists permanently,
send instrument for repair
Depending on the reason for the fault and the measures taken, the
stepsdescribedinchapter"Set up"mayhavetobecarriedoutagain.
8.3 Instrument repair
If a repair is necessary, please proceed as follows:
You can download a return form (23 KB) from our Internet homepage
www.vega.comunder:"Downloads - Forms and certicates - Repair
form".
By doing this you help us carry out the repair quickly and without hav-
ing to call back for needed information.
Printandlloutoneformperinstrument
Clean the instrument and pack it damage-proof
Attach the completed form and, if need be, also a safety data
sheet outside on the packaging
Please ask the agency serving you for the address of your return
shipment.Yo ucanndtherespectiveagencyonourwebsitewww.
vega.comunder:"Company - VEGA worldwide".
Note:
If a battery-operated VEGA instrument is returned for repair, the bat-
tery/power pack must be removed. Please send only the instrument,
not the battery/power pack.