8Maintenanceandfaultrectication
PLICSMOBILE T61 • External GSM/GPRS radio unit
37700-EN-130228
8 Maintenanceandfaultrectication
8.1 Maintenance
If the instrument is used properly, no special maintenance is required
in normal operation.
8.2 Rectify faults
The operator of the system is responsible for taking suitable meas-
ures to rectify faults.
A maximum of reliability is ensured. Nevertheless, faults can occur
during operation. These may be caused by the following, e.g.:
•
Measured value from sensor not correct
•
Voltage supply
•
Interference on the cables
Therstmeasurestobetakenaretochecktheinput/outputsignals
as well as to evaluate the error messages via the DTM. The procedure
is described below. In many cases, the causes can be determined in
thiswayandfaultscanberectied.
Should these measures not be successful, please call in urgent cases
the VEGA service hotline under the phone no. +49 1805 858550.
The hotline is available to you 7 days a week round-the-clock. Since
weoerthisserviceworld-wide,thesupportisonlyavailableinthe
English language. The service is free of charge, only the standard
telephone costs will be charged.
Error code Cause Rectication
No indication
of the signal
strength
No GSM network
available
– Check network availability via mobile
phone
E008 Sensor not found – Check connection of the sensor
E013 Sensor signals
failure, no valid
measured value
– Check sensor parameter adjustment
– Send sensor for repair
E030 Sensor in boot phase
value not valid
– Check sensor parameter adjustment
E034 EEPROM CRC error – Switchtheinstrumentoandon
– Carry out a reset
– Send instrument for repair
E035 ROM CRC error – Switchtheinstrumentoandon
– Carry out a reset
– Send instrument for repair
E036 Instrument software
not executable (dur-
ing software update
or after failed update)
– Waituntilsoftwareupdateisnished
– Carry out another software update
Reaction when malfunc-
tions occur
Failure reasons
Faultrectication