Operating InstructionsVEGAPULS 624 … 20 mA/HART - two-wire, approval according to LPR radiostandardDocument ID:41718Radar
Transport must be carried out under consideration of the notes on thetransport packaging. Nonobservance of these instructions can causedamage to the d
VEGADIS 62 is suitable for measured value indication and adjustmentof sensors with HART protocol. It is looped into the 4 … 20 mA/HARTsignal cable.You
4 Mounting4.1 General instructionsWith instruments with threaded process fitting, suitable tools must beapplied for tightening the hexagon.Warning:The
Note:The radar sensor with rinsing air connection or antenna extension isprovided with a notch on the antenna socket. This notch must bealigned with t
The position of the polarisation plane is marked on the process fittingof the instrument.11Fig. 3: Orientation of the polarisation plane with threaded
> 200 mm(7.87")Fig. 5: Mounting of the radar sensor on round vessel topsIn vessels with conical bottom it can b e advantageous to mount thesen
Fig. 7: Mounting of the radar sensor with inflowing mediumThe socket piece should be dimensioned in such a way that theantenna end protrudes at least 1
dhmax.d1½"50 mm/2"80 mm/3"100 mm/4"150 mm/6"200 mm250 mm300 mm500 mm800 mmhmax.Fig. 9: Deviating socket dimensionsTip:The ins
Fig. 11: Cover smooth profiles with deflectorsIf there are agitators in the vessel, a false signal suppression should becarried out with the agitators i
When using a surge pipe in a vessel, influences from vesselinstallations and turbulences can be excluded. Under these prereq-uisites, the measurement o
Contents1 About this document1.1 Function. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .41.2 Target group . . . . . . . . . . . .
l The 100 % point must be below the upper vent hole and theantenna edgel The 0 % point is the end of the surge pipel For the parameter adjustment, sel
0 %100 %125436Fig. 14: Configuration bypass1 Radar sensor2 Marking of the polarisation plane3 Instrument flange4 Distance sensor reference plane to uppe
Constructional requirements on the bypass pipe:l Material metal, smoother inner tubel In case of an extremely rough tube inner surface, use an inserte
3 ... 4 hmax≥ 50 mm90°231hmaxdmin≥ 2 mm x hmax90°42 3Fig. 16: Flow measurement with rectangular flume: dmin.= min. distance of thesensor (see chapter &
23 ... 4 x hmax90°hmaxdB1Fig. 17: Flow measurement with Khafagi-Venturi flume: d = Min. distance tosensor; hmax.= max. filling of the flume; B = tightest
5 Connecting to power supply5.1 Preparing the connectionAlways keep in mind the following safety instructions:l Connect only in the complete absence o
If screened cable is necessary, connect the cable screen on both endsto ground potential. In the sensor, the screen must be connecteddirectly to the i
5 Insert the cable into the sensor through the cable entryFig. 18: Connection steps 5 and 66 Insert the wire ends into the terminals according to the
512+( )(-)6 7 84...20mA2341Fig. 19: Electronics and connection compartment, single chamber housing1 Voltage supply, signal output2 For indicating and
4...20mA2312+( )(-)1Fig. 21: Connection compartment, double chamber housing1 Voltage supply, signal output2 For indicating and adjustment module or i
8 Set up with other systems8.1 DD adjustment programs . . . . . . . . . . . . . . . . . . . . .538.2 Communicator 375, 475 . . . . . . . . . . . . . .
5.5 Wiring plan, Ex-d double chamber housing5 6 7812+( )(-)3124...20mAFig. 23: Electronics compartment, double chamber housing1 Internal connection t
Contact pin Colour connectioncable in the sensorTerminal, electronicsmodulePin 1 Brown 5Pin 2 White 6Pin 3 Blue 7Pin 4 Black 85.6 Wiring plan - versio
6 Set up with the indicating and adjustmentmodule6.1 Insert indicating and adjustment moduleThe indicating and adjustment module can be inserted into
1 2Fig. 28: Insertion of the indicating and adjustment module with double chamberhousing1 In the electronics compartment2 In the connection compartmen
6.2 Adjustment system12Fig. 29: Indicating and adjustment elements1 LC display2 Adjustment keysl [OK] key:- Move to the me nu overview- Confirm selecte
6.3 Parameter adjustmentThrough the parameter adjustment the instrument is adapted to theapplication conditions. The parameter adjustment is carried o
Through this selection, the sensor is adapted perfectly to the productand measurement reliability, particularly in products with poorreflective propert
l Product speed: slow filling and emptyingl Process/measurement conditions:- Condensation- Smooth product surface- Max. requirement to the measurement
Standpipe:l Product speed: very fast filling and emptyingl Vessel:- Vent hole- Joins like flanges, weld joints- Shifting of the running time in the tube
- Ice and condensation on the antenna possible- Spiders and insect nestle in the antennas- Floating material and animals sporadically on the watersurf
1 About this document1.1 FunctionThis operating instructions manual provides all the information youneed for mounting, connection and setup as well as
l Vessel of metal: weld jointsl Process/measurement conditions:- Filling too close to the sensor- System noise with completely empty silo increased- A
Through this sel ection, the sensor is adapted optimally to theapplication or the location and measurement reliability under thevarious basic conditio
The real product level during this adjustment is not important, becausethe min./max. adjustment is always carried out without changing theproduct leve
3 Set the requested percentage value with [+] and save with [OK].The cursor jumps now to the distance value.4 Enter the appropriate distance value in
A comparison of echo curve and false signal suppression allows amore accurate conclusion on measurement reliability. The selectedcurve is continuously
1 Select the menu item "Additional settins" with [->] and confirm with[OK]. With [->] you have to select the menu item "False sign
The menu item "Extend" is used to extend an already created falsesignal suppression. This is useful if a false signal suppression wascarried
Delivery status: Restoring the parameter settings at the time ofshipment from the factory incl. the order-specific settings. A createdfalse signal supp
Menu Menu item Default valueLock operation ReleasedDisplay Language Like orderDisplayed value DistanceDisplay unit mScaling size VolumelScaling 0.00 l
VEGAPULS 62 • 4 … 20 mA/HART - two-wire, approval according to LPR radio standard 496 Set up with the indicating and adjustment module41718-EN-120301
2 For your safety2.1 Authorised personnelAll operations described in this operating instructions manual must becarried out only by trained specialist
7 Setup with PACTware7.1 Connect the PC312Fig. 30: Connection of the PC directly to the sensor via the interface adapter1 USB cable to the PC2 Interfa
Note:With power supply units with integrated HART resistance (internalresistance approx. 250 Ω), an additional external resistance is notnecessary. Th
All device DTMs are available as a free-of-charge standard versionand as a full version that must be purchased. In the standard version,all functions
8 Set up with other systems8.1 DD adjustment programsDevice descriptions as Enhanced Device Description (EDD) areavailable for DD adjustment programs
9 Diagnosis, Asset Management and service9.1 MaintenanceIf the device is used correctly, no maintenance is required in normaloperation.9.2 Diagnosis m
Further echo curves: Up to 10 echo curves can be stored in a ringbuffer in this memory section.Additional echo curves are stored via:l PC with PACTware
Maintenance: Due to external influences, the instrument function islimited. The measurement is affected, but the measured value is stillvalid. Plan in m
CodeText mes-sageCause RectificationF125Unpermissibleelectronicstemperaturel Temperature of the elect-ronics in the non-specifiedsectionl Check ambient
CodeText mes-sageCause RectificationS600Unpermissibleelectronicstemperaturel Temperature of the elect-ronics in the non-specifiedsectionl Check ambient
CodeText mes-sageCause RectificationM505No echo avai-lablel Level echo can no longerbe detectedl Clean the antennal Use a more suitable an-tenna/sensor
Depending on the instrument version, the emitting frequencies are inthe C or K band range. The low emitting frequencies are far below theinternational
The below tables show typical examples of application-relevantmeasurement errors with liquids. The measurement errors aredifferentiated according to th
Fault description Error pattern Cause Rectification3. Measured valuejumps towards 100 %Leveltime0l Due to the process, the amp-litude of the product ec
Fault description Error pattern Cause Rectification8. Measured valuejumps towards 100 %during fillingLeveltime0l Due to strong turbulence andfoam genera
Depending on the reason for the fault and the measures taken, thesteps described in chapter "Setup" must be carried out again or mustbe chec
l Go under "www.vega.com/downloads" to "Software"l Select under "plics sensors and instruments", "Firmware updates&
l Please ask the agency serving you for the address of your returnshipment. You can find the competent agency on our websitewww.vega.com.VEGAPULS 62 •
10 Dismounting10.1 Dismounting stepsWarning:Before dismounting, be aware of dangerous process conditions suchas e.g. pressure in the vessel, high temp
11 Supplement11.1 Technical dataGeneral data316L corresponds to 1.4404 or 1.4435Materials, wetted parts- Process fitting 316L, Hastelloy C22, Monell Al
Input variableMeasured variable The measured quantity is the distance betweenprocess fitting of the sensor and product surface.The reference plane is t
- QV (Fourth Value) Scaled measured valueResolution, digital < 1 mm (0.039 in)Accuracy (according to DIN EN 60770-1)Process reference conditions ac
Not included in the CE confirmity declaration are the countriesimplementing this radio standard at a later date: Finland, Luxembourg,Slovenia and Hunga
Specifications apply also to the current outputTemperature drift - Current output ±0.03 %/10 K relating to the 16 mA span max.±0.3 %Deviation on the cu
Tracking speed of the measuring windowmax.1 m/minBeam angle3)- Horn antenna ø 75 mm (2.953 in) 10°- Horn antenna ø 95 mm (3.74 in) 8°Emitted HF power4
- Antenna impedance cone PEEK -1 … 100 bar (-100 … 10000 kPa/-14.5 … 1450 ps ig)- Antenna impedance cone ceramic -1 … 160 bar (-100 … 16000 kPa/-14.5
or:l 1 x closing cap M20 x 1.5; 1 x blind stopperM20 x 1 .5or:l 1 x closing cap ½ NPT, 1 x blind plug ½ NPTor:l 1 x plug (depending on the version), 1
- Colour - standard PE Black- Colour - standard PUR Blue- Colour - Ex-version BlueIndicating and adjustment moduleVoltage supply and data transmission
Permissible residual ripple - Non-Ex, Ex-ia instrument- for 9.6 V< UN< 14 V ≤ 0.7 Veff(16 … 400 Hz)- for 18 V< UN< 36 V ≤ 1.0 Veff(16 … 400
76 VEGAPULS 62 • 4 … 20 mA/HART - two-wire, approval according to LPR radio standard11 Supplement41718-EN-120301
11.2 Radio astronomy stationsThe following table shows the geographic position of the radio astronomy stations in Europe:Country Name of the Station G
11.3 DimensionsThe following dimensional drawings represent only an extract of thepossible versions. Detailed dimensional drawings can be downloadedon
Aluminium housing in protection rating IP 66/IP 68 (1 bar)120 mm (4.72")~ 105 mm (4.13")116 mm (4.57")~ 150 mm (5.91")ø 84 mm (3.3
3 Product description3.1 StructureThe type label contains the most important data for identification anduse of the instrument:21131415161211536478910Fi
Stainless steel housing in protection rating IP 66/IP 68, 1 bar117 mm (4.61")~ 103 mm(4.06")ø 77 mm(3.03")120 mm (4.72")~ 105 mm (
VEGAPULS 62, horn antenna in threaded versionmminch38 mm (1.50")22 mm (0.87")22 mm (0.87")SW 46 mm(1.81")ø2.95" 8.50"16.
VEGAPULS 62, horn antenna in flange versionxy21165 mm (6.5")60 mm(2.36")mminchø2.95" 8.50"16.93"ø3.74"yxø75 216ø95 430yxF
VEGAPULS 62, horn antenna in flange version 450 °Cyxinchø 3.74" 8.50"11.30"ø 2.95"mmyxø 75 216ø 95 287yx260 mm (10.24")Fig. 57
VEGAPULS 62, horn antenna and swivelling holdermminch20 mm(0.79")20 mm(0.79")ø2.95" 8.50"16.93"ø3.74"x y xy140 mm(1.58&q
11.4 Industrial property rightsVEGA product lines are global protected by industrial property rights.Further information see http://www.vega.com.Only
INDEXAAccessory- External indicating and adjustment unit 10-11- External radio unit 11- Flanges 11- Indicating and adjustment module 10- Interface ada
Polarisation plane 13Potential equalisation 26RReflection properties 36Repair 64Replacement parts- Antenna impedance cone 11- Electronics module 11Rese
VEGA Grieshaber KGAm Hohenstein 11377761 SchiltachGermanyPhone +49 7836 50-0Fax +49 7836 50-201E-mail: [email protected] date:ISO 9
This operating instructions manual applies to the following instrumentversions:l Hardware from 1.0.0l Software from 4 .4.0The scope of delivery encomp
Comments to this Manuals