Vega VEGAFLEX 63 Spacer Cable-ø 4 mm _ Rod-ø 10 mm User Manual

Browse online or download User Manual for Equipment Vega VEGAFLEX 63 Spacer Cable-ø 4 mm _ Rod-ø 10 mm. VEGA VEGAFLEX 63 Spacer Cable-ø 4 mm _ Rod-ø 10 mm User Manual

  • Download
  • Add to my manuals
  • Print
  • Page
    / 8
  • Table of contents
  • BOOKMARKS
  • Rated. / 5. Based on customer reviews
Page view 0
Supple
mentary instructions
Spacer
for VEGAFLEX 63
Cable-ø 4 mm / R
od-ø 10 mm
Document ID:
33028
G
uided Microwave
Page view 0
1 2 3 4 5 6 7 8

Summary of Contents

Page 1 - Cable-ø 4 mm / R

Supplementary instructionsSpacerfor VEGAFLEX 63Cable-ø 4 mm / Rod-ø 10 mmDocument ID:33028Guided Microwave

Page 2 - 1 Produc

1 Product descriptionIf VEGAFLEX sensors are used in standpipes or bypass tubes, contactto the tube wall must be avoided.The spacer fixes the probe in

Page 3 - 33028-EN-100527

2 Mounting1Fig. 1: Mountingof the spacer1 Spacer1 Measure the diameter of the standpipe or bypass tube and adaptthe spacers accordingly. The markings

Page 4 - 3.1 Technical data

3 Supplement3.1 Technical dataGeneral dataMaterial- Spacer (4-leg) PEEKDiameter ø 95 mm (3.7 in) markings ø 75 mm andø 48 mm (3 in and 1.9 in)Product

Page 5

3.2 DimensionsSpacer for cable-ø 4 mm (0.16 in)5 mm(0,2")5 mm(13/64")5,5 mm(0.22")ø 48 mm (1.89")ø 75 mm (2.95")ø 95 mm (3.74

Page 6

Spacer for rod-ø 10 mm (0.39 in)5,5 mm(0.22")ø 48 mm (1.89")ø 75 mm (2.95")5 mm(0.2")ø10 mm(0.39")11 mm(0.43")ø 95 mm (3

Page 7

Spacer • Cable-ø 4 mm / Rod-ø 10 mm 73 Supplement33028-EN-100527

Page 8 - ISO 9001

VEGA Grieshaber KGAm Hohenstein 11377761 SchiltachGermanyPhone +49 7836 50-0Fax +49 7836 50-201E-mail: [email protected] date:ISO 9