Operating InstructionsVEGACONNECT 4 with connectionboxInterface converter USB - HART/I²CDocument ID:32628Signalconditioning instrumentsand communicati
213OPENTWISTUSBLOCKFig. 2: Electricalconnection1 USB 1.1 or 2.0 connection of the PC2 USB connection cable (in the scope of delivery)3 VEGACONNECT 4 w
1 32OPENTWISTUSBLOCKFig. 3: ConnectionpossibilitiesCommunication via I²C protocoll Use of VEGACONNECT 4 in all plics®sensors through directinstallatio
Note:The connection of Profibus PA or Foundation Fieldbus instruments isalways carried out via the I²C interface on the sensor. A directconnection of V
5 Connection examples5.1 Connection via I²C interfaceIn hazardous Ex areas, the HART or I²C cable of VEGACONNECT 4can be connected to an Ex approved s
OPENTWISTUSBLOCK1 2Fig. 6: Connection VEGADIS 61 via I²C interface1 I²C bus (Com.) interface2 I²C connectioncableIf the sensors in the HART or Profibus
12OPENTWISTUSBLOCKFig. 8: Connection VEGAMET/VEGASCAN via I²C interface1 I²C bus (Com.) interface2 I²C connectioncableNote:Communication with the sens
VEGAMET 381on1212%132OPENTWISTUSBLOCKFig. 9: Connectionplics®series via HART1 HART sensor2 Connection cable with 2 mm pins3 VEGAMET 381 or VEGADIS 371
6 Setup6.1 OperationUse and adjustment are described in the operating instructions manualas well as in the online help of PACTware and the VEGA-DTMs a
7 Maintenance and fault rectification7.1 MaintenanceWhen used as directed in normal operation, VEGACONNECT 4 iscompletely maintenance free.7.2 Instrume
8 Dismount8.1 Dismounting stepsSince VEGACONNECT 4 is a mobile instrument, it is not necessary totake note of separate dismounting steps.8.2 DisposalT
Contents1 About this document1.1 Function. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .31.2 Target group . . . . . . . . . . . .
9 Supplement9.1 Technical dataElectrical dataVoltage supply from USB interface 5 VMax. power consumption 500 mWGalvanic separation between l HART - US
9.2 Dimensions162mm (6.38") 52 mm(2.05")80mm (3.15")Fig. 12: Dimensions VEGACONNECT 4VEGACONNECT 4 withconnection box • Interface con
9.3 Industrial property rightsVEGA product lines are global protected by industrial property rights.Further information see http://www.vega.com.Only i
INDEX44 … 20 mA 7CConnection 12DDriver 9DTM 7, 17DTM Collection 9FFoundation Fieldbus 7, 12HHART 7- Modem 7- Resistor 11-12, 15II²C interface 11Interf
VEGA Grieshaber KGAm Hohenstein 11377761 SchiltachGermanyPhone +49 7836 50-0Fax +49 7836 50-201E-mail: [email protected] date:ISO 9
1 About this document1.1 FunctionThis operating instructions manual provides all the information youneed for mounting, connection and setup as well as
2 For your safety2.1 Authorised personnelAll operations described in this operating instructions manual must becarried out only by trained specialist
2.5 Safetylabel on the instrumentThe safety approval markings and safety tips on the device must beobserved.2.6 CE conformityThis device fulfills the l
3 Product description3.1 ConfigurationThe scope of delivery encompasses:l VEGACONNECT 4 interface converterl Connection box with two connection cablesl
l Technical datal Article numbers, documentationThe serial number allows you to access the delivery data of theinstrument viawww.vega.com, "VEGA
3.3 OperationThe adjustment is carried out via a Windows PC with a parameteradjustment software such as PACTware with respective DTM. Thereare no adju
4 Connection4.1 Connection to the PCNote:First of all you should install the driver before connecting VEGA-CONNECT 4 to the PC.An USB interface (1.1 o