Operating InstructionsPressure transmitter with ceramic measuring cellVEGABAR 82Modbus and Levelmaster protocolDocument ID: 46294
103 Product descriptionVEGABAR 82 • Modbus and Levelmaster protocol46294-EN-13110712453Fig. 3: Recessed installation of the CERTEC® measuring cell1 M
113 Product descriptionVEGABAR 82 • Modbus and Levelmaster protocol46294-EN-13110712345Fig. 5: Front-ush installation of the CERTEC® measuring cell w
123 Product descriptionVEGABAR 82 • Modbus and Levelmaster protocol46294-EN-131107The delivery must be checked for completeness and possible transit d
133 Product descriptionVEGABAR 82 • Modbus and Levelmaster protocol46294-EN-131107Youcanndadditionalinformationinthesupplementaryinstructions
144 MountingVEGABAR 82 • Modbus and Levelmaster protocol46294-EN-1311074 Mounting4.1 General instructions to use the instrumentMake sure that all par
154 MountingVEGABAR 82 • Modbus and Levelmaster protocol46294-EN-13110712Fig. 7: Temperature ranges1 Process temperature2 Ambient temperature4.2 Ve
164 MountingVEGABAR 82 • Modbus and Levelmaster protocol46294-EN-1311071 2 3666 6654Fig. 8: Position of the lter element - non-Ex, Ex-ia and Ex-d-ia
174 MountingVEGABAR 82 • Modbus and Levelmaster protocol46294-EN-131107213434Fig. 9: Position of the lter element - Ex-d version1 Single chamber hou
184 MountingVEGABAR 82 • Modbus and Levelmaster protocol46294-EN-131107• Atmospheric pressure compensation (with relative pressure meas-uring ranges)1
194 MountingVEGABAR 82 • Modbus and Levelmaster protocol46294-EN-13110714321234Fig. 13: Measurement setup with process pressure measurement of gases i
2ContentsVEGABAR 82 • Modbus and Levelmaster protocol46294-EN-131107Contents1 About this document1.1 Function ...
204 MountingVEGABAR 82 • Modbus and Levelmaster protocol46294-EN-131107312Fig. 14: Measurement setup for process pressure measurement of liquids in pi
214 MountingVEGABAR 82 • Modbus and Levelmaster protocol46294-EN-1311074.5 External housing12345Fig. 16: Setup process assembly, external housing1 P
225 Connecting to power supply and bus systemVEGABAR 82 • Modbus and Levelmaster protocol46294-EN-1311075 Connecting to power supply and bus system5.
235 Connecting to power supply and bus systemVEGABAR 82 • Modbus and Levelmaster protocol46294-EN-1311075.2 ConnectionThe voltage supply and signal ou
245 Connecting to power supply and bus systemVEGABAR 82 • Modbus and Levelmaster protocol46294-EN-1311077. Connect the screen to the internal ground
255 Connecting to power supply and bus systemVEGABAR 82 • Modbus and Levelmaster protocol46294-EN-131107Information:The connection of an external disp
265 Connecting to power supply and bus systemVEGABAR 82 • Modbus and Levelmaster protocol46294-EN-131107123512( )+(-)6784...20mAFig. 23: Electronics
275 Connecting to power supply and bus systemVEGABAR 82 • Modbus and Levelmaster protocol46294-EN-131107++power supplyMODBUSD0D1IS GNDUSB124 313452o
286 Set up with the display and adjustment moduleVEGABAR 82 • Modbus and Levelmaster protocol46294-EN-1311076 Set up with the display and adjustment
296 Set up with the display and adjustment moduleVEGABAR 82 • Modbus and Levelmaster protocol46294-EN-1311076.2 Adjustment system12Fig. 27: Display a
3ContentsVEGABAR 82 • Modbus and Levelmaster protocol46294-EN-1311078.5 Calculation of total deviation (according to DIN 16086) ...
306 Set up with the display and adjustment moduleVEGABAR 82 • Modbus and Levelmaster protocol46294-EN-131107Carry out the following steps in the below
316 Set up with the display and adjustment moduleVEGABAR 82 • Modbus and Levelmaster protocol46294-EN-1311076. Min. adjustmentIn this menu item you ca
326 Set up with the display and adjustment moduleVEGABAR 82 • Modbus and Levelmaster protocol46294-EN-131107• Special characters +, -, /, -In this men
336 Set up with the display and adjustment moduleVEGABAR 82 • Modbus and Levelmaster protocol46294-EN-131107Temperature unit:Enter the requested param
346 Set up with the display and adjustment moduleVEGABAR 82 • Modbus and Levelmaster protocol46294-EN-13110721100%0%5 m(196.9")Fig. 28: Parameter
356 Set up with the display and adjustment moduleVEGABAR 82 • Modbus and Levelmaster protocol46294-EN-1311074. Move with [ESC] and [->] to the spa
366 Set up with the display and adjustment moduleVEGABAR 82 • Modbus and Levelmaster protocol46294-EN-1311075. Save settings with [OK] and move with
376 Set up with the display and adjustment moduleVEGABAR 82 • Modbus and Levelmaster protocol46294-EN-131107user especially when adjusting the switchi
386 Set up with the display and adjustment moduleVEGABAR 82 • Modbus and Levelmaster protocol46294-EN-131107The default setting for the display value
396 Set up with the display and adjustment moduleVEGABAR 82 • Modbus and Levelmaster protocol46294-EN-131107Selecttherequestedsimulationsizeanda
41 About this documentVEGABAR 82 • Modbus and Levelmaster protocol46294-EN-1311071 About this document1.1 FunctionThis operating instructions manual
406 Set up with the display and adjustment moduleVEGABAR 82 • Modbus and Levelmaster protocol46294-EN-131107Reset - SetupMenu item Parameter Default v
416 Set up with the display and adjustment moduleVEGABAR 82 • Modbus and Levelmaster protocol46294-EN-131107Menu item Parameter Default valueSimulatio
426 Set up with the display and adjustment moduleVEGABAR 82 • Modbus and Levelmaster protocol46294-EN-131107In this menu item you gain access to the p
436 Set up with the display and adjustment moduleVEGABAR 82 • Modbus and Levelmaster protocol46294-EN-1311076.5 Saving the parameter adjustment dataW
447 Setting up sensor and Modbus interface with PACTwareVEGABAR 82 • Modbus and Levelmaster protocol46294-EN-1311077 Setting up sensor and Modbus int
457 Setting up sensor and Modbus interface with PACTwareVEGABAR 82 • Modbus and Levelmaster protocol46294-EN-1311071Fig. 30: Connecting the PC via USB
467 Setting up sensor and Modbus interface with PACTwareVEGABAR 82 • Modbus and Levelmaster protocol46294-EN-131107Note:To ensure that all instrument
477 Setting up sensor and Modbus interface with PACTwareVEGABAR 82 • Modbus and Levelmaster protocol46294-EN-1311077.3 Saving the parameter adjustmen
488 Diagnosis, asset management and serviceVEGABAR 82 • Modbus and Levelmaster protocol46294-EN-1311078 Diagnosis, asset management and service8.1 Ma
498 Diagnosis, asset management and serviceVEGABAR 82 • Modbus and Levelmaster protocol46294-EN-131107the following tables there are more detailed err
52 For your safetyVEGABAR 82 • Modbus and Levelmaster protocol46294-EN-1311072 For your safety2.1 Authorised personnelAll operations described in th
508 Diagnosis, asset management and serviceVEGABAR 82 • Modbus and Levelmaster protocol46294-EN-131107CodeText mes-sageCause RecticationF013no measur
518 Diagnosis, asset management and serviceVEGABAR 82 • Modbus and Levelmaster protocol46294-EN-131107CodeText mes-sageCause RecticationF265Measureme
528 Diagnosis, asset management and serviceVEGABAR 82 • Modbus and Levelmaster protocol46294-EN-131107CodeText mes-sageCause RecticationM501Error in
538 Diagnosis, asset management and serviceVEGABAR 82 • Modbus and Levelmaster protocol46294-EN-1311078.5 Calculation of total deviation (according t
548 Diagnosis, asset management and serviceVEGABAR 82 • Modbus and Levelmaster protocol46294-EN-131107Caution:The exchange may only be carried out in
558 Diagnosis, asset management and serviceVEGABAR 82 • Modbus and Levelmaster protocol46294-EN-131107In Ex applications, only instruments and electro
569 DismountingVEGABAR 82 • Modbus and Levelmaster protocol46294-EN-1311079 Dismounting9.1 Dismounting stepsWarning:Before dismounting, be aware of d
5710 SupplementVEGABAR 82 • Modbus and Levelmaster protocol46294-EN-13110710 Supplement10.1 Technical dataMaterials and weightsMaterials, wetted part
5810 SupplementVEGABAR 82 • Modbus and Levelmaster protocol46294-EN-131107Input variableAdjustmentAdjustment range of the min./max. adjustment relatin
5910 SupplementVEGABAR 82 • Modbus and Levelmaster protocol46294-EN-131107Nominal range Overload capacity, max. pressureOverload capacity, min. pres-s
62 For your safetyVEGABAR 82 • Modbus and Levelmaster protocol46294-EN-131107radiated emissions must be taken into account, as is usual with a class A
6010 SupplementVEGABAR 82 • Modbus and Levelmaster protocol46294-EN-131107Nominal range Overload capacity, max. pressureOverload capacity, min. pres-s
6110 SupplementVEGABAR 82 • Modbus and Levelmaster protocol46294-EN-131107Additional output parameter - temperatureProcessingismadeviaoutputsigna
6210 SupplementVEGABAR 82 • Modbus and Levelmaster protocol46294-EN-131107Medium or ambient tem-perature TAccuracy class 0.05 %, 0.1 %Accuracy class 0
6310 SupplementVEGABAR 82 • Modbus and Levelmaster protocol46294-EN-131107Long-term drift of the zero signal - Version climate-compensatedNominal meas
6410 SupplementVEGABAR 82 • Modbus and Levelmaster protocol46294-EN-131107Measuring cell seal Product temperature - standard versionProduct temperatur
6510 SupplementVEGABAR 82 • Modbus and Levelmaster protocol46294-EN-131107Electromechanical data - version IP 66/IP 68 (1 bar)Connection cable Ʋ Cong
6610 SupplementVEGABAR 82 • Modbus and Levelmaster protocol46294-EN-131107 Ʋ mounted into the housing without coverIP 40Materials Ʋ Housing ABS Ʋ Insp
6710 SupplementVEGABAR 82 • Modbus and Levelmaster protocol46294-EN-131107Housing material Version IP-protection class NEMA protectionAluminium Single
6810 SupplementVEGABAR 82 • Modbus and Levelmaster protocol46294-EN-131107413 3 32++power supplyMODBUSD0D1IS GNDUSB++power supplyMODBUSD0D1IS GNDU
6910 SupplementVEGABAR 82 • Modbus and Levelmaster protocol46294-EN-131107GeneralcongurationofthehostThedataexchangewithstatusandvariables
73 Product descriptionVEGABAR 82 • Modbus and Levelmaster protocol46294-EN-1311073 Product description3.1 CongurationThenameplatecontainsthemos
7010 SupplementVEGABAR 82 • Modbus and Levelmaster protocol46294-EN-131107Register Name Register Number Type NoteSV Unit 108 DWord Unit CodeSV 110 Sec
7110 SupplementVEGABAR 82 • Modbus and Levelmaster protocol46294-EN-131107Register Name Register Number Type NotePV 2202 DWord Primary Variable in Byt
7210 SupplementVEGABAR 82 • Modbus and Levelmaster protocol46294-EN-131107Parameter Length Code/DataStart Address 2 Bytes 2*NRegister Value N*2 Bytes
7310 SupplementVEGABAR 82 • Modbus and Levelmaster protocol46294-EN-131107Parameter Length Code/DataSub Function Code 2 BytesData N*2 Bytes DataRespon
7410 SupplementVEGABAR 82 • Modbus and Levelmaster protocol46294-EN-131107Response:Parameter Length Code/DataFunction Code 1 Byte 0x11Byte Number 1 By
7510 SupplementVEGABAR 82 • Modbus and Levelmaster protocol46294-EN-131107Parameter CongurableValues Default ValueBaud Rate 1200, 2400, 4800, 9600,
7610 SupplementVEGABAR 82 • Modbus and Levelmaster protocol46294-EN-131107Response:Parameter Length Code/DataReport Level (and Temperature) 6 characte
7710 SupplementVEGABAR 82 • Modbus and Levelmaster protocol46294-EN-131107Set Receive to Transmit DelayRequest:Parameter Length Code/DataSet Receive t
7810 SupplementVEGABAR 82 • Modbus and Levelmaster protocol46294-EN-13110710.6 CongurationoftypicalModbushostsFisher ROC 809Wiring plan++power
7910 SupplementVEGABAR 82 • Modbus and Levelmaster protocol46294-EN-131107ABB Total FlowWiring plan++power supplyMODBUSD0D1IS GNDUSB1 213452offon(
83 Product descriptionVEGABAR 82 • Modbus and Levelmaster protocol46294-EN-131107This operating instructions manual applies to the following instrumen
8010 SupplementVEGABAR 82 • Modbus and Levelmaster protocol46294-EN-131107Thermo Electron AutopilotWiring plan++power supplyMODBUSD0D1IS GNDUSB12313
8110 SupplementVEGABAR 82 • Modbus and Levelmaster protocol46294-EN-131107Bristol ControlWave MicroWiring plan++power supplyMODBUSD0D1IS GNDUSB12313
8210 SupplementVEGABAR 82 • Modbus and Levelmaster protocol46294-EN-131107ScadaPackWiring plan++power supplyMODBUSD0D1IS GNDUSB13213455213496782offo
8310 SupplementVEGABAR 82 • Modbus and Levelmaster protocol46294-EN-131107dimensional drawings can be downloaded at www.vega.com under "Downloads
8410 SupplementVEGABAR 82 • Modbus and Levelmaster protocol46294-EN-131107External housing with IP 68 version1234582 mm(3.23")80 mm(3.15")10
8510 SupplementVEGABAR 82 • Modbus and Levelmaster protocol46294-EN-131107VEGABAR82,threadedttingnotfront-ushDD DLDUC2DQDIG½ G¼ G½ G¼ SW 27 mm(
8610 SupplementVEGABAR 82 • Modbus and Levelmaster protocol46294-EN-131107VEGABAR82,threadedttingfront-ushC1C322 mm(0.87")44 mm(1.73")
8710 SupplementVEGABAR 82 • Modbus and Levelmaster protocol46294-EN-131107VEGABAR82,hygienicttingARAAESFR EZ/E1 E350 mm(1.97")51 mm(2.01"
8810 SupplementVEGABAR 82 • Modbus and Levelmaster protocol46294-EN-131107VEGABAR82,angeconnection3.46"4.02"5.43" 0.12" 0.12&
8910 SupplementVEGABAR 82 • Modbus and Levelmaster protocol46294-EN-131107VEGABAR82,extensionttingF1 F2AF/AGF9ø 32...60 mm(1.26...2.36")ø 48.
93 Product descriptionVEGABAR 82 • Modbus and Levelmaster protocol46294-EN-131107• VacuumMeasured products are gases, vapours and liquids.Dependingon
9010 SupplementVEGABAR 82 • Modbus and Levelmaster protocol46294-EN-13110710.8 Industrial property rightsVEGA product lines are global protected by i
91INDEXVEGABAR 82 • Modbus and Levelmaster protocol46294-EN-131107INDEXAAdjust Date/Time 39Adjustment – Level 35, 36 – Process pressure 34, 35 – Sy