Operating InstructionsVEGAFLEX 66-20 … +250 °CProfibus PADocument ID:34171Guided Microwave

Page 2

The delivery must be checked for completeness and possible transitdamage immediately at receipt. Ascertained transit damage orconcealed defects must b

Page 3

4 Mounting4.1 General instructionsMake sure that all parts of the instrument exposed to the process, inparticular the sensor element, process seal and

Page 4 - 1.3 Symbolism used

Fig. 2: Measuresagainst moisture penetrationThe reference plane for the measuring range of the sensors is thesealing surface of the thread or flange.Ke

Page 5 - 2.3 Warning about misuse

keep in mind that measurement down to the tip of the probe is notpossible. The exact value of the min. distance (lower dead band) isstated in chapter

Page 6

Excessive system vibration or shocks, e.g. caused by agitators orturbulence in the vessel (e.g. from fluidisation) can cause the coaxprobe of VEGAFLEX

Page 7 - 3 Produc

1Fig. 5: Positionof the spacer1 SpacerNote:Measurement in a standpipe is not recommended for very adhesiveproducts.If there is a danger of the probe t

Page 8 - 3.2 Principle of operation

1 2 Fig. 6: Vesselwith temperature insulation1 Temperature insulation2 Ambient temperature on the housing16 VEGAFLEX 66 • Profibus PA4 Mounting34171-EN

Page 9 - 3.3 Operation

5 Connecting to power supply5.1 Preparing the connectionAlways keep in mind the following safety instructions:l Connect only in the complete absence o

Page 10 - 34171-EN-100427

Insystems with potential equalisation, connect the cable screendirectly to ground potential at the power supply unit, in the connectionbox and at the

Page 11 - 4.1 General instructions

7 Insertthe wire ends into the open terminals according to the wiringplanFig. 7: Connectionsteps 6 and 78 Press down the opening levers of the termina

Page 12 - 4.2 Mounting instructions

Contents1 About this document1.1 Function. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .41.2 Target group . . . . . . . . . . . .

Page 13

I²CDisplay1 2 5 6 7 83412Fig. 8: Electronicsand connection compartment with single chamber housing1 Plug connector for VEGACONNECT (I²C interface)2 Sp

Page 14

132Display1 2 5 6 7 8I²CFig. 10: Electronicscompartment, double chamber housing1 Plug connector for VEGACONNECT (I²C interface)2 Internal connection c

Page 15

I2C11 2Fig. 12: Wiringplan with double chamber housing1 Voltage supply/Signal output5.5 Wiring plan - version IP 66/IP 68, 1 bar+-12Fig. 13: Wireassig

Page 16

6 Setup with the indicating and adjustmentmodule PLICSCOM6.1 Short descriptionThe indicating and adjustment module is used for measured valuedisplay,

Page 17 - 5.1 Preparing the connection

Fig. 14: Insertindicating and adjustment moduleNote:Ifyou intend to retrofit the instrument with an indicating and adjustmentmodule for continuous meas

Page 18 - 5.2 Connection procedure

6.3 Adjustment system1.1231Fig. 15: Indicatingand adjustment elements1 LC display2 Indication of the menu item number3 Adjustment keysl [OK] key:- Mov

Page 19

6.4 SetupstepsAfter VEGAFLEX 66 is connected to voltage supply or after voltagerecurrence, the instrument carries out a self-check for approx. 30secon

Page 20

Start your parameter adjustment with the following menu items of thebasic adjustment:Proceed as follows:1 Move from the measured value display to the

Page 21

Eachproduct has different reflective properties. In addition, there arevarious interfering factors which have to be taken into account:agitated product

Page 22

percentage and vessel volume. Byactivating the appropriate curve,the volume percentage of the vessel is displayed correctly. If thevolume should not b

Page 23 - 6.1 Short description

8 Maintenance and fault rectification8.1 Maintenance . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .388.2 Rectify malfunctions . . . . . .

Page 24

Gating out of false sig nalsChange now?Proceed as follows:1 Move from the measured value display to the main menu bypushing [OK].2 Select the menu ite

Page 25

Copy sensor dataCopy sensor data?Basic adjustmentIf the function "Reset" is carried out, the sensor resets the values of thefollowing menu i

Page 26 - 6.4 Setup

Additional adjustment and diagnosis options such as e.g. scaling,simulation or trend curve presentation are shown in the following menuschematic. You

Page 27

6.5 MenuschematicBasic adjustment1▶Basic adjustmentDisplayDiagnosticsServiceInfo1.1Sensor address0001.2Min. adjustment0.00 %=10.000 m(d)8.000 m(d)1.3M

Page 28

Service4Basic adjustmentDisplayDiagnostics▶ServiceInfo4.1Sensor5 mRod4.2ApplicationLiquidStandard(DK ≥ 2)4.3Gating out of false sig nalsChange now?4.4

Page 29 - Linearisation curve

6.6 Savingthe parameter adjustment dataIt is recommended noting the adjusted data, e.g. in this operatinginstructions manual and archive them afterwar

Page 30 - Change now?

7 Setup with PACTware and other adjustmentprograms7.1 Connecting the PC312Fig. 16: Connectionof the PC via VEGACONNECT directly to the sensor1 USB cab

Page 31

Necessary components:l VEGAFLEX 66l PC with PACTware and suitable VEGA DTMl VEGACONNECTl Power supply unit or processing system7.2 Parameter adjustmen

Page 32

8 Maintenance and fault rectification8.1 MaintenanceWhen used as directed in normal operation, VEGAFLEX 66 iscompletely maintenance free.8.2 Rectify ma

Page 33 - 6.5 Menu

Error Cause RemovalIn Simatic S7 themeasuredvalue isalways presentedas 0Only four bytesare consistentlyloaded in the PLCUse function component SFC 14

Page 34

1 About this document1.1 FunctionThis operating instructions manual provides all the information youneed for mounting, connection and setup as well as

Page 35 - 6.6 Saving

Statusaccor-ding to NE 107Error code Text message Cause/RectificationE017 Adjustment span too small Adjustment not within the specifica-tion. Carry out

Page 36 - 7.1 Connecting the PC

l FX-E.60PX (X = withoutapprovals)l FX-E.60PA (A = approvals CA, DA according to product list)l FX-E.60PC (C = approvals XM, CX, CM, CK, GX according

Page 37

l Print and fill out one form per instrumentl Clean the instrument and pack it damage-proofl Attach the completed form and, if need be, also a safety d

Page 38 - 8.2 Rectify malfunctions

9 Dismounting9.1 Dismounting stepsWarning:Before dismounting, be aware of dangerous process conditions suchas e.g. pressure in the vessel, high temper

Page 39

10 Supplement10.1 Technical dataGeneral dataMaterial 316L corresponds to 1.4404 or 1.4435Materials, wetted parts- Process fitting - coax version 316L a

Page 40

- Rod: ø 6 mm (0.236 in) approx. 220 g/m (2.365 oz/ft)- Cable: ø 2 mm (0.079 in) approx. 20 g/m (0.215 oz/ft)- Cable: ø 4 mm (0.157 in) approx. 80 g/m

Page 41 - 8.5 Instrument repair

24315Fig. 18: Measuringrange of VEGAFLEX 66 - coax version1 Reference plane2 Probe length3 Measuring range4 Upper dead band (see diagrams under Accura

Page 42

2431245315Fig. 19: Measuringranges of the VEGAFLEX 66 - rod and cable version1 Reference plane2 Probe length3 Measuring range (default setting refers

Page 43 - 9 Dismou

- Relative humidity 45 … 75 %- Air pressure +860 … +1060 mbar/+86 … +106 kPa(+12.5 … +15.4 psig)Installation reference conditions- Min. distance to in

Page 44 - 10 Supplem

L0-3mm (-0.118")3mm (0.118")15mm(0.591")-10mm (-0.394")0,25 m(9.843")0,08 m(3.15")0,02 m(0.787")1Fig. 20: Devia

Page 45

2 For your safety2.1 Authorised personnelAll operations described in this operating instructions manual must becarried out only by trained specialist

Page 46

L0-3mm (-0.118")3mm (0.118")15mm(0.591")-15mm (-0.591")0,25 m(9.843")0,08 m(3.149")0,25 m(9.843")1Fig. 22: Devi

Page 47

L0-3mm (-0.118")3mm (0.118")-50mm (1.969")5mm (0.197")0,15 m (5.906")0,03 m (1.181")0,02 m(0.787")1Fig. 24: Dev

Page 48

Process conditionsProcess pressure -1 … +100 bar/-100 … +10000 kPa (--14.5 … +1450 psig), depending on the processfittingProcess temperature -20 … +250

Page 49

l 1 xclosing cap ½ NPT, 1 x blind stopper½ NPT , 1 x blind stopper M16 x 1.5 oroptionally 1 x plug M12 x 1 for VEGADIS 61or:l 1 x plug (depending on t

Page 50

Voltage supplyOperating voltage- Non-Ex instrument 9 … 32 V DC- EEx-ia instrument 9 … 24 V DC- EEx-d instrument 16 … 32 V DCOperating voltage with lig

Page 51

10.2 Profibus PAInstrument master fileThe instrument master file (GSD) contains the characteristic data of the Profibus PA instrument.These data are, e.g.

Page 52

Module of the PA sensorsFor the cyclic data traffic, VEGAFLEX 66 provides the following modules:l AI (PA-OUT)- PA-OUT value of the FB1 after scalingl Ad

Page 53

l Free PlaceByte-No.1 2 3 4 5StatusStatus(FB1)IEEE-754-Floating point valuePA- OUT(FB1)FormatValueDataformat of the output signalByte4Byte3Byte2Byte1B

Page 54

StatuscodeDescriptionaccording to Pro-fibus standardPossible cause0x0C bad - sensor failure l Hardware errorl Converter errorl Leakage pulse errorl Tri

Page 55 - 10.2 Profibus P

10.3 DimensionsThe following dimensional drawings represent only an extract of thepossible versions. Detailed dimensional drawings can be downloadedon

Page 56 - - Additi

2.5 Safetylabel on the instrumentThe safety approval markings and safety tips on the device must beobserved.2.6 CE conformityThis device fulfills the l

Page 57

Aluminium housing in protection rating IP 66/IP 68, 1 bar120 mm (4.72")~ 105 mm (4.13")116 mm (4.57")~ 150 mm (5.91")ø 84 mm (3.31

Page 58

VEGAFLEX 66 - cable, rodversion (-20 … +250 °C/-4 … +482 °F)SW41 (G1 / 1 NPT)SW46 (G1½ / 1½ NPT)G1 / 1 NPT,G1½ / 1½ NPTL140 mm (5.51")20 mm (0.79

Page 59 - 10.3 Dim

VEGAFLEX 66 - coax version (-20 … +250 °C/-4 … +482 °F)LG¾ / ¾ NPT,G1 / 1 NPT,G1½ / 1½ NPT140 mm (5.51")20 mm (0.79")27 mm (1.06")ø 21,

Page 60

10.4 Industrial property rightsVEGA product lines are global protected by industrial property rights.Further information see http://www.vega.com.Only

Page 61 - G1½ / 1½ NPT

VEGA Grieshaber KGAm Hohenstein 11377761 SchiltachGermanyPhone +49 7836 50-0Fax +49 7836 50-201E-mail: [email protected] date:ISO 9

Page 62 - (0.84")

3 Product description3.1 StructureThe scope of delivery encompasses:l VEGAFLEX 66 level sensorl Documentation- this operating instructions manual- Ope

Page 63 - 10.5 Trademark

123Fig. 1: VEGAFLEX 66 incable version with plastic housing1 Housing cover with integrated indicating and adjustment module (optional)2 Housing with e

Page 64 - ISO 9001

Power supply via the Profibus DP/PA segment coupler or VEGALOG571 EP cards. A two-wire cable according to Profibus specificationserves a s carrier of bot