Operating InstructionsPressure transmitter with CERTEC® measuring cellVEGABAR 524 … 20 mA/HARTDocument ID: 36717
103 Product descriptionVEGABAR 52 • 4 … 20 mA/HART36717-EN-130321The delivery must be checked for completeness and possible transit damage immediately
113 Product descriptionVEGABAR 52 • 4 … 20 mA/HART36717-EN-130321Youcanndadditionalinformationinthesupplementaryinstructionsmanual"Fla
124 MountingVEGABAR 52 • 4 … 20 mA/HART36717-EN-1303214 Mounting4.1 General instructionsMake sure that all parts of the instrument coming in direct c
134 MountingVEGABAR 52 • 4 … 20 mA/HART36717-EN-13032121212112Fig. 4: Position of the lter element1 Filter element2 Blind plugCaution:Duetothel
144 MountingVEGABAR 52 • 4 … 20 mA/HART36717-EN-1303214.2 Mounting stepsFormountingVEGABAR52,aweldedsocketisrequired.Youcanndthesecompo
154 MountingVEGABAR 52 • 4 … 20 mA/HART36717-EN-130321Tip:Mount the wall mounting plate so that the cable entry of the socket housing points downward.
165 Connecting to power supplyVEGABAR 52 • 4 … 20 mA/HART36717-EN-1303215 Connecting to power supply5.1 Preparing the connectionAlways keep in mind
175 Connecting to power supplyVEGABAR 52 • 4 … 20 mA/HART36717-EN-130321housing.Thiswouldinuencethestabilityoftheconnectionandthetightness
185 Connecting to power supplyVEGABAR 52 • 4 … 20 mA/HART36717-EN-1303218. Press down the opening levers of the terminals, you will hear the terminal
195 Connecting to power supplyVEGABAR 52 • 4 … 20 mA/HART36717-EN-130321321Fig. 8: Components of the external housing1 Screws2 Wall mounting plate3
2ContentsVEGABAR 52 • 4 … 20 mA/HART36717-EN-130321Contents1 About this document1.1 Function ...
205 Connecting to power supplyVEGABAR 52 • 4 … 20 mA/HART36717-EN-130321I²CDisplay12 56783412Fig. 9: Electronics and connection compartment, single ch
215 Connecting to power supplyVEGABAR 52 • 4 … 20 mA/HART36717-EN-130321132Display12 5678I2CFig. 11: Electronics compartment, double chamber housing1
225 Connecting to power supplyVEGABAR 52 • 4 … 20 mA/HART36717-EN-130321I2C112Fig. 13: Wiring plan, double chamber housing1 Voltage supply, signal ou
235 Connecting to power supplyVEGABAR 52 • 4 … 20 mA/HART36717-EN-1303211212Fig. 15: Connection compartment, Ex-d double chamber housing1 Spring-load
245 Connecting to power supplyVEGABAR 52 • 4 … 20 mA/HART36717-EN-1303215.7 Wiring plan, external housing with version IP 68 (25 bar)Fig. 18: VEGABAR
255 Connecting to power supplyVEGABAR 52 • 4 … 20 mA/HART36717-EN-1303211234634125Fig. 20: Connection of the sensor in the housing socket1 Brown2 Bl
265 Connecting to power supplyVEGABAR 52 • 4 … 20 mA/HART36717-EN-130321Then the corresponding current is outputted to the cable (the value correspond
276 Set up with the display and adjustment module PLICSCOMVEGABAR 52 • 4 … 20 mA/HART36717-EN-1303216 Set up with the display and adjustment module P
286 Set up with the display and adjustment module PLICSCOMVEGABAR 52 • 4 … 20 mA/HART36717-EN-130321Fig. 22: Insert display and adjustment moduleNote:
296 Set up with the display and adjustment module PLICSCOMVEGABAR 52 • 4 … 20 mA/HART36717-EN-130321 – Save value• [->] key to select: – Menu chang
3ContentsVEGABAR 52 • 4 … 20 mA/HART36717-EN-1303218.1 Maintain ...
306 Set up with the display and adjustment module PLICSCOMVEGABAR 52 • 4 … 20 mA/HART36717-EN-130321Information:The steps 1, 3 and 4 are not necessary
316 Set up with the display and adjustment module PLICSCOMVEGABAR 52 • 4 … 20 mA/HART36717-EN-1303214. Conrmwith[OK],thesubmenu"Density un
326 Set up with the display and adjustment module PLICSCOMVEGABAR 52 • 4 … 20 mA/HART36717-EN-1303214. Set the requested mbar value with [+] and [-&g
336 Set up with the display and adjustment module PLICSCOMVEGABAR 52 • 4 … 20 mA/HART36717-EN-130321Inthemenuitems"zero"and"span&
346 Set up with the display and adjustment module PLICSCOMVEGABAR 52 • 4 … 20 mA/HART36717-EN-130321UnitUnitofmeasurementbar▼Temperature unit°C▼3.
356 Set up with the display and adjustment module PLICSCOMVEGABAR 52 • 4 … 20 mA/HART36717-EN-130321Information:Thezeroadjustmentshiftsthevalueo
366 Set up with the display and adjustment module PLICSCOMVEGABAR 52 • 4 … 20 mA/HART36717-EN-130321Caution:Note the following if the VEGABAR 52 with
376 Set up with the display and adjustment module PLICSCOMVEGABAR 52 • 4 … 20 mA/HART36717-EN-130321Basic adjustmentThe"Reset""Basic
386 Set up with the display and adjustment module PLICSCOMVEGABAR 52 • 4 … 20 mA/HART36717-EN-130321itemsintheoperatinginstructionsmanual"D
396 Set up with the display and adjustment module PLICSCOMVEGABAR 52 • 4 … 20 mA/HART36717-EN-130321DiagnosticsBasic adjustment 3Display▶DiagnosticsSe
41 About this documentVEGABAR 52 • 4 … 20 mA/HART36717-EN-1303211 About this document1.1 FunctionThis operating instructions manual provides all the
407 Set up with PACTware and other adjustment programsVEGABAR 52 • 4 … 20 mA/HART36717-EN-1303217 Set up with PACTware and other adjustment programs7
417 Set up with PACTware and other adjustment programsVEGABAR 52 • 4 … 20 mA/HART36717-EN-13032112 43OPENTWISTUSBLOCKFig. 26: Connecting the PC via HA
427 Set up with PACTware and other adjustment programsVEGABAR 52 • 4 … 20 mA/HART36717-EN-1303217.3 Parameter adjustment with AMS™ and PDMFor VEGA se
438MaintenanceandfaultrecticationVEGABAR 52 • 4 … 20 mA/HART36717-EN-1303218 Maintenanceandfaultrectication8.1 MaintainIf the instrument is
448MaintenanceandfaultrecticationVEGABAR 52 • 4 … 20 mA/HART36717-EN-130321Error code Cause Rectication4 … 20 mA signal not stableLeveluctuati
458MaintenanceandfaultrecticationVEGABAR 52 • 4 … 20 mA/HART36717-EN-130321Fperf=√((FT)2 + (FKl)2 + (Fa)2)With:• Ftotal: Total deviation• Fperf
468MaintenanceandfaultrecticationVEGABAR 52 • 4 … 20 mA/HART36717-EN-130321Total deviation analogue output signal absolute:Ftotal = 0.365 % x 8 b
478MaintenanceandfaultrecticationVEGABAR 52 • 4 … 20 mA/HART36717-EN-130321data and displays the actual hardware and software version of the sens
489 DismountingVEGABAR 52 • 4 … 20 mA/HART36717-EN-1303219 Dismounting9.1 Dismounting stepsWarning:Before dismounting, be aware of dangerous process
4910 SupplementVEGABAR 52 • 4 … 20 mA/HART36717-EN-13032110 Supplement10.1 Technical dataGeneral dataParameter, pressure Gauge pressure, absolute pre
52 For your safetyVEGABAR 52 • 4 … 20 mA/HART36717-EN-1303212 For your safety2.1 Authorised personnelAll operations described in this operating inst
5010 SupplementVEGABAR 52 • 4 … 20 mA/HART36717-EN-130321 Ʋ Connection cable with IP 68 1 bar versionPEWeight approx.0.8 … 8 kg (1.764 … 17.64 lbs), d
5110 SupplementVEGABAR 52 • 4 … 20 mA/HART36717-EN-130321Accuracy Ʋ in the range of 0 … +100 °C (+32 … +212 °F)±3 K Ʋ in the range of -50 … 0 °C (-58
5210 SupplementVEGABAR 52 • 4 … 20 mA/HART36717-EN-130321Nominalrange Overload capacity, max. pressureOverload capacity, min. pres-sure-0.2 … +0.2 ba
5310 SupplementVEGABAR 52 • 4 … 20 mA/HART36717-EN-130321Nominalrange Overload capacity, max. pressureOverload capacity, min. pres-sure0 … 14.5 psi 5
5410 SupplementVEGABAR 52 • 4 … 20 mA/HART36717-EN-130321Applies to the digitalsignaloutput(HART,ProbusPA,FoundationFieldbus)aswellastoan
5510 SupplementVEGABAR 52 • 4 … 20 mA/HART36717-EN-130321 Ʋ Version IP 66/IP 68 (1 bar), connec-tion cable PE-20 … +60 °C (-4 … +140 °F)Process condit
5610 SupplementVEGABAR 52 • 4 … 20 mA/HART36717-EN-130321 Ʋ Single chamber housing – 1 x cable gland M20 x 1.5 (cable: ø 5 … 9 mm), 1 x blind plug M2
5710 SupplementVEGABAR 52 • 4 … 20 mA/HART36717-EN-130321Electromechanical data - version IP 68Connection cable between IP 68 instrument and external
5810 SupplementVEGABAR 52 • 4 … 20 mA/HART36717-EN-130321 Ʋ 100Hz…10kHz Uss < 10 mVLoad see diagram100075050025012 181614 20 22 24 26 28 30 32
5910 SupplementVEGABAR 52 • 4 … 20 mA/HART36717-EN-130321For that reason the associated approval documents of these instruments have to be carefully n
62 For your safetyVEGABAR 52 • 4 … 20 mA/HART36717-EN-1303212.6 CE conformityThisdevicefulllsthelegalrequirementsoftheapplicableECguide-li
6010 SupplementVEGABAR 52 • 4 … 20 mA/HART36717-EN-130321Aluminium housing in protection rating IP 66/IP 68 (1 bar)120 mm (4.72")~ 105 mm (4.13&q
6110 SupplementVEGABAR 52 • 4 … 20 mA/HART36717-EN-130321Stainless steel housing in protection rating IP 66/IP 68 (1 bar)117 mm (4.61")~ 103 mm(4
6210 SupplementVEGABAR 52 • 4 … 20 mA/HART36717-EN-13032112368 mm(2.68")68 mm(2.68")92 mm(3.62")42mm(1.65")40mm(1.57")110 mm
6310 SupplementVEGABAR 52 • 4 … 20 mA/HART36717-EN-130321VEGABAR52,threadedtting1GI GM/GNGV/GCGBGSGL/GUG½AG¼AG½AG¼AG½A½"NPTM20x1,5SW 27 mm(1
6410 SupplementVEGABAR 52 • 4 … 20 mA/HART36717-EN-130321VEGABAR52,threadedtting2GFGEGWGGSW46SW55G1 ½ AG2 AG1 ½ A50 mm(1.97")55 mm(2.17"
6510 SupplementVEGABAR 52 • 4 … 20 mA/HART36717-EN-130321VEGABAR52,hygienictting1SW4651 mm(2 1/64")51 mm(2 1/64")51 mm(2 1/64")60
6610 SupplementVEGABAR 52 • 4 … 20 mA/HART36717-EN-130321VEGABAR52,hygienictting2SW4650 mm(1 31/32")51 mm(2 1/64")60 mm(2 23/64")5
6710 SupplementVEGABAR 52 • 4 … 20 mA/HART36717-EN-130321VEGABAR52,angeconnection3.46"4.02"5.43" 0.12" 0.12" 0.12"
6810 SupplementVEGABAR 52 • 4 … 20 mA/HART36717-EN-130321VEGABAR52,threadedttingforpaperindustryM 44 x 1,25SW 46 mm (1.81")M 56 x 1,25 M 4
6910 SupplementVEGABAR 52 • 4 … 20 mA/HART36717-EN-130321VEGABAR52,angeconnectionforpaperindustry1FGTRTSEW158 mm(6.22")40 mm(1.58")ø
73 Product descriptionVEGABAR 52 • 4 … 20 mA/HART36717-EN-1303213 Product description3.1 CongurationThe scope of delivery encompasses:• VEGABAR 52
7010 SupplementVEGABAR 52 • 4 … 20 mA/HART36717-EN-130321VEGABAR52,angeconnectionforpaperindustry243 mm(1.69")ø 48 mm(1.89")23 mm(0.
7110 SupplementVEGABAR 52 • 4 … 20 mA/HART36717-EN-13032110.3 Industrial property rightsVEGA product lines are global protected by industrial propert
72INDEXVEGABAR 52 • 4 … 20 mA/HART36717-EN-130321INDEXAAccessories – Display and adjustment module 10 – External display and adjustment unit 10 – Ex
73NotesVEGABAR 52 • 4 … 20 mA/HART36717-EN-130321
74NotesVEGABAR 52 • 4 … 20 mA/HART36717-EN-130321
75NotesVEGABAR 52 • 4 … 20 mA/HART36717-EN-130321
Printing date:VEGA Grieshaber KGAm Hohenstein 11377761 SchiltachGermany36717-EN-130321All statements concerning scope of delivery, application, practi
83 Product descriptionVEGABAR 52 • 4 … 20 mA/HART36717-EN-1303212113105436789111214Fig. 2: Layout of the type label (example)1 Instrument type2 Prod
93 Product descriptionVEGABAR 52 • 4 … 20 mA/HART36717-EN-130321The CERTEC® measuring cell is also equipped with a temperature sensor. The temperature