Vega VEGACONNECT User Manual Page 3

  • Download
  • Add to my manuals
  • Print
  • Page
    / 4
  • Table of contents
  • BOOKMARKS
  • Rated. / 5. Based on customer reviews
Page view 2
VEGACONNECT
3
Technical Information
2.3 Sensor in conjunction with VEGALOG 571
1
15
16
sensors
VEGACONNECT
5
4
3
2
1
9
8
7
6
VBUS
2 mm-plug
RS 232
D-Sub-plug
9-pole
GND
RxD
TxD
146
1
2
Sensor
Ex-area
not-Ex-area
e.g. VEGAMET 514 V
CONNECT-
sockets in the
front plate
connection
line
2.5 Ex-sensors with VEGAMET 509 V, 512 V,
514 V or VEGAMET 515 V
Connection points Parameter adjustment
1 VEGAMET
Sensor
2 Sensor
3 Electrical connection
4 Technical data
Power supply from PC/LAPTOP via
RS 232
Power consumption max. 8 mA
Voltage max. 15 V
Transmission rate 9.600 baud
Data format 8 bit (ASCII-Code)
Parity even
Start / Stop 1 bit each
Galvanic separation between VBUS and RS 232
Line length RS 232 approx. 30 cm
VBUS approx. 150 cm
Permissible
ambient temperature -20°C … +60°C
Storage and
transport temperature -20°C … +70°C
Housing material shock-resistant Polystyrol
upper part light grey
lower part agate
Housing dimensions W x H x D
= approx. 100 x 52 x 30 mm
Weight
- with connection line approx. 200 g
connection line
CONNECT CONNECT CONNECT CONNECT CONNECT
Connection points Parameter adjustment
1 … 15 respective sensor
16 Sensor 1 … 15
2.4 Compact and dust Ex-sensors
2
1
Sensor
Power supply
connection line to
indicator
CONNECT
Connection points Parameter adjustment
1 Sensor
2 Sensor
Note
A digital signal (VBUS) is additionally modulated to the
0/4 … 20 mA-output. Acc. to the connected output (e.g.
indicator) this signal is damped by the internal resistor of
the output, if it is smaller than 100 . For analog inputs of
PLC-systems this resistor is generally 100 , so that no
damping is caused. In case of a low impedance indicator a
load resistor of R = 100 must be connected in the
0/4 … 20 mA-line and VEGACONNECT must be
connected to this resistor.
CONNECT-sockets on
the sensor
CONNECT-sockets on
the sensor
EV-card CPU-card
to sensor
+
-
R = 100
VEGACONNECT
Page view 2
1 2 3 4

Comments to this Manuals

No comments