Vega VEGABAR 41 Profibus PA User Manual Page 25

  • Download
  • Add to my manuals
  • Print
  • Page
    / 36
  • Table of contents
  • BOOKMARKS
  • Rated. / 5. Based on customer reviews
Page view 24
VEGABAR 41 (Profibus PA) 25
Here you can choose in which unit the adjust-
ment should be made (mbar, psi, kPa). Then
you enter the pressure values corresponding
to 0 % and 100 %. Then click to OK. The
adjustment is now finished.
Click in the window "Adjustment“ to Zero/
Span.
Then in the window "Zero/Span adjustment“,
you have to choose the type of adjustment. If
you want to carry out the "Adjustment without
taking the actual product into account“ (dry
adjustment), click to no, then to OK.
If you want to carry out the "Adjustment
taking the actual product into account“ (live
adjustment), click in this window to yes, then
to OK.
Choose the measurement loop in which you
want to make the parameter adjustment and
click to OK.
From the window "Instrument data parameter
adjustment“ you reach all submenus of the
sensor. To get fimiliar with the available
options, try out the buttons one after the
other. You always return to this window. To
carry out the adjustment, click the button
Adjustment.
Sensor parameter adjustment – Adjust-
ment
Click to Instrument data, then to Parameter
adjustment.