Operating InstructionsRadiation-based sensor for continuous level and interface measurementSOLITRAC 314 … 20 mA/HART - four-wireDocument ID: 40090
103 Product descriptionSOLITRAC 31 • 4 … 20 mA/HART - four-wire40090-EN-1311193.5 Corresponding source containerA radioactive isotope in a suitable s
113 Product descriptionSOLITRAC 31 • 4 … 20 mA/HART - four-wire40090-EN-131119Control areas are areas in which the local dose rate exceeds a certain v
124 MountingSOLITRAC 31 • 4 … 20 mA/HART - four-wire40090-EN-1311194 Mounting4.1 General instructionsThe source container is part of the measuring sy
134 MountingSOLITRAC 31 • 4 … 20 mA/HART - four-wire40090-EN-1311194.2 Mounting instructionsNote:Along with the planning, our specialists will analys
144 MountingSOLITRAC 31 • 4 … 20 mA/HART - four-wire40090-EN-131119To measure the level in very high vessels, multiple instruments can be cascaded.Cas
154 MountingSOLITRAC 31 • 4 … 20 mA/HART - four-wire40090-EN-131119On vessels with temperature insulation, the sensor and the source container should
165 Connecting to power supplySOLITRAC 31 • 4 … 20 mA/HART - four-wire40090-EN-1311195 Connecting to power supply5.1 Preparing the connectionAlways
175 Connecting to power supplySOLITRAC 31 • 4 … 20 mA/HART - four-wire40090-EN-131119If potential equalisation currents are expected, the connection o
185 Connecting to power supplySOLITRAC 31 • 4 … 20 mA/HART - four-wire40090-EN-1311191Fig. 6: Connection steps 4 and 51 Locking of the terminal block
195 Connecting to power supplySOLITRAC 31 • 4 … 20 mA/HART - four-wire40090-EN-131119Information:The terminal blocks are pluggable and can be detached
2ContentsSOLITRAC 31 • 4 … 20 mA/HART - four-wire40090-EN-131119Contents1 About this document1.1 Function ...
205 Connecting to power supplySOLITRAC 31 • 4 … 20 mA/HART - four-wire40090-EN-131119567821Fig. 8: Adjustment and connection compartment with non-Ex i
215 Connecting to power supplySOLITRAC 31 • 4 … 20 mA/HART - four-wire40090-EN-131119567823412+( )(-)1Fig. 10: Adjustment and connection compartment
225 Connecting to power supplySOLITRAC 31 • 4 … 20 mA/HART - four-wire40090-EN-1311195.3 Connection - Level detectionNon-Ex instruments and instrumen
235 Connecting to power supplySOLITRAC 31 • 4 … 20 mA/HART - four-wire40090-EN-131119instructions are part of the scope of delivery and come with the
245 Connecting to power supplySOLITRAC 31 • 4 … 20 mA/HART - four-wire40090-EN-1311195.4 Connection - SummationTo measure the level in very high vess
255 Connecting to power supplySOLITRAC 31 • 4 … 20 mA/HART - four-wire40090-EN-131119112456 910112211122122 1920 18 17 14 1316 15 12112456 91011221112
265 Connecting to power supplySOLITRAC 31 • 4 … 20 mA/HART - four-wire40090-EN-131119Information:For example, a radial connection would be also possib
276 Set up with the display and adjustment moduleSOLITRAC 31 • 4 … 20 mA/HART - four-wire40090-EN-1311196 Set up with the display and adjustment modu
286 Set up with the display and adjustment moduleSOLITRAC 31 • 4 … 20 mA/HART - four-wire40090-EN-1311196.2 Adjustment system12Fig. 17: Display and a
296 Set up with the display and adjustment moduleSOLITRAC 31 • 4 … 20 mA/HART - four-wire40090-EN-131119Information:Inthisoperatinginstructionsman
3ContentsSOLITRAC 31 • 4 … 20 mA/HART - four-wire40090-EN-1311199.5 Software update ...
306 Set up with the display and adjustment moduleSOLITRAC 31 • 4 … 20 mA/HART - four-wire40090-EN-131119For this purpose, check which isotope is integ
316 Set up with the display and adjustment moduleSOLITRAC 31 • 4 … 20 mA/HART - four-wire40090-EN-131119In this menu item you can enter the measuring
326 Set up with the display and adjustment moduleSOLITRAC 31 • 4 … 20 mA/HART - four-wire40090-EN-131119Select "Show table"todisplayande
336 Set up with the display and adjustment moduleSOLITRAC 31 • 4 … 20 mA/HART - four-wire40090-EN-131119Note:Ifyoucannotllthevesselduringthea
346 Set up with the display and adjustment moduleSOLITRAC 31 • 4 … 20 mA/HART - four-wire40090-EN-131119With the setting "Automatic", the in
356 Set up with the display and adjustment moduleSOLITRAC 31 • 4 … 20 mA/HART - four-wire40090-EN-131119You can choose whether the sensor should outpu
366 Set up with the display and adjustment moduleSOLITRAC 31 • 4 … 20 mA/HART - four-wire40090-EN-131119DisplayWith this parameter you can change the
376 Set up with the display and adjustment moduleSOLITRAC 31 • 4 … 20 mA/HART - four-wire40090-EN-131119Pulse rate of the sensorProcess valueCurrent o
386 Set up with the display and adjustment moduleSOLITRAC 31 • 4 … 20 mA/HART - four-wire40090-EN-131119Peak values of measured value: Resetting of th
396 Set up with the display and adjustment moduleSOLITRAC 31 • 4 … 20 mA/HART - four-wire40090-EN-131119Menu Menu item Default valueAdditional adjustm
41 About this documentSOLITRAC 31 • 4 … 20 mA/HART - four-wire40090-EN-1311191 About this document1.1 FunctionThis operating instructions manual prov
406 Set up with the display and adjustment moduleSOLITRAC 31 • 4 … 20 mA/HART - four-wire40090-EN-131119The address setting (MGC) on the Slave instrum
416 Set up with the display and adjustment moduleSOLITRAC 31 • 4 … 20 mA/HART - four-wire40090-EN-131119SetupIn this menu item you can assign an unamb
426 Set up with the display and adjustment moduleSOLITRAC 31 • 4 … 20 mA/HART - four-wire40090-EN-131119instrument.Order-specicsettingsremainbut
436 Set up with the display and adjustment moduleSOLITRAC 31 • 4 … 20 mA/HART - four-wire40090-EN-131119Menu Menu item Default valueAdditional adjustm
446 Set up with the display and adjustment moduleSOLITRAC 31 • 4 … 20 mA/HART - four-wire40090-EN-131119parameters to ensure optimum adjustment of the
456 Set up with the display and adjustment moduleSOLITRAC 31 • 4 … 20 mA/HART - four-wire40090-EN-131119the component of the pulse rate originating fr
466 Set up with the display and adjustment moduleSOLITRAC 31 • 4 … 20 mA/HART - four-wire40090-EN-131119You can enter the value for "Adjustment u
476 Set up with the display and adjustment moduleSOLITRAC 31 • 4 … 20 mA/HART - four-wire40090-EN-131119You can let the adjustment point be determined
486 Set up with the display and adjustment moduleSOLITRAC 31 • 4 … 20 mA/HART - four-wire40090-EN-131119Inthismenuitemyoucanalsodenetheswitc
496 Set up with the display and adjustment moduleSOLITRAC 31 • 4 … 20 mA/HART - four-wire40090-EN-131119The peak value function holds the max. and min
52 For your safetySOLITRAC 31 • 4 … 20 mA/HART - four-wire40090-EN-1311192 For your safety2.1 Authorised personnelAll operations described in this o
506 Set up with the display and adjustment moduleSOLITRAC 31 • 4 … 20 mA/HART - four-wire40090-EN-131119Information:The simulation is automatically te
516 Set up with the display and adjustment moduleSOLITRAC 31 • 4 … 20 mA/HART - four-wire40090-EN-131119Menu Menu item Default valueSetup Measurement
526 Set up with the display and adjustment moduleSOLITRAC 31 • 4 … 20 mA/HART - four-wire40090-EN-131119• Instrument name - shows instrument name and
537 Setup with PACTwareSOLITRAC 31 • 4 … 20 mA/HART - four-wire40090-EN-1311197 Setup with PACTware7.1 Connect the PC213Fig. 18: Connection of the P
547 Setup with PACTwareSOLITRAC 31 • 4 … 20 mA/HART - four-wire40090-EN-131119• SOLITRAC 31• PCwithPACTwareandsuitableVEGADTM• VEGACONNECT4• HA
557 Setup with PACTwareSOLITRAC 31 • 4 … 20 mA/HART - four-wire40090-EN-131119Fig. 20: Example of a DTM viewAll device DTMs are available as a free-of
568 Set up with other systemsSOLITRAC 31 • 4 … 20 mA/HART - four-wire40090-EN-1311198 Set up with other systems8.1 DD adjustment programsDevice desc
579 Diagnostics and serviceSOLITRAC 31 • 4 … 20 mA/HART - four-wire40090-EN-1311199 Diagnostics and service9.1 MaintenanceIf the device is used corre
589 Diagnostics and serviceSOLITRAC 31 • 4 … 20 mA/HART - four-wire40090-EN-131119Maintenance:Duetoexternalinuences,theinstrumentfunctionisl
599 Diagnostics and serviceSOLITRAC 31 • 4 … 20 mA/HART - four-wire40090-EN-131119CodeText mes-sageCause RecticationF036Faulty pro-gram memory – Erro
62 For your safetySOLITRAC 31 • 4 … 20 mA/HART - four-wire40090-EN-1311192.5 CE conformityThedevicefulllsthelegalrequirementsoftheapplicable
609 Diagnostics and serviceSOLITRAC 31 • 4 … 20 mA/HART - four-wire40090-EN-131119CodeText mes-sageCause RecticationF120Filter time error – Faulty or
619 Diagnostics and serviceSOLITRAC 31 • 4 … 20 mA/HART - four-wire40090-EN-131119CodeText mes-sageCause RecticationS017Accuracy outside the specica
629 Diagnostics and serviceSOLITRAC 31 • 4 … 20 mA/HART - four-wire40090-EN-131119Error Cause Rectication4 … 20 mA signal missingElectrical connec-ti
639 Diagnostics and serviceSOLITRAC 31 • 4 … 20 mA/HART - four-wire40090-EN-131119Sinceweoerthisserviceworldwide,thesupportisprovidedinEng
649 Diagnostics and serviceSOLITRAC 31 • 4 … 20 mA/HART - four-wire40090-EN-131119Youcanndarepairformaswellasdetailedinformationonhowto
6510 DismountingSOLITRAC 31 • 4 … 20 mA/HART - four-wire40090-EN-13111910 Dismounting10.1 Dismounting stepsWarning:Before dismounting, be aware of da
6611 SupplementSOLITRAC 31 • 4 … 20 mA/HART - four-wire40090-EN-13111911 Supplement11.1 Technical dataGeneral data316L corresponds to 1.4404 or 1.443
6711 SupplementSOLITRAC 31 • 4 … 20 mA/HART - four-wire40090-EN-13111912LFig. 22: Data of the input variable1 Min. level (red marking line)2 Max. le
6811 SupplementSOLITRAC 31 • 4 … 20 mA/HART - four-wire40090-EN-131119Output variable - Level measurementOutput signals 4 … 20 mA/HART - active; 4 … 2
6911 SupplementSOLITRAC 31 • 4 … 20 mA/HART - four-wire40090-EN-131119 Ʋ Max. 3 A AC, 1 A DCBreaking capacity Ʋ Min. 50 mW Ʋ Max. 750VAAC,40WDCIf
73 Product descriptionSOLITRAC 31 • 4 … 20 mA/HART - four-wire40090-EN-1311193 Product description3.1 CongurationThenameplatecontainsthemostim
7011 SupplementSOLITRAC 31 • 4 … 20 mA/HART - four-wire40090-EN-131119Process conditionsFortheprocessconditions,pleasealsonotethespecications
7111 SupplementSOLITRAC 31 • 4 … 20 mA/HART - four-wire40090-EN-131119Measurement electronics temeratureResolution 1 °C (1.8 °F)Accuracy ±1 °C (1.8 °F
7211 SupplementSOLITRAC 31 • 4 … 20 mA/HART - four-wire40090-EN-131119Aluminium and stainless steel housing242,5 mm (9.55")54,5 mm(2.15")9 m
7311 SupplementSOLITRAC 31 • 4 … 20 mA/HART - four-wire40090-EN-131119SOLITRAC 3122 mm(0.87")36 mm(1.42")ø 76,2 mm(3")L242,5 mm (9.55&q
7411 SupplementSOLITRAC 31 • 4 … 20 mA/HART - four-wire40090-EN-131119SOLITRAC 31 - Mounting example220 mm(8.66")70 mm(2.76")101,6 mm(4.00&q
7511 SupplementSOLITRAC 31 • 4 … 20 mA/HART - four-wire40090-EN-13111911.3 Industrial property rightsVEGA product lines are global protected by indus
76INDEXSOLITRAC 31 • 4 … 20 mA/HART - four-wire40090-EN-131119INDEXAAccessories – External indicating unit 9Adjustment 31 – System 28Adjustment dat
77INDEXSOLITRAC 31 • 4 … 20 mA/HART - four-wire40090-EN-131119Summation Slave 39TTime 37, 50Type plate 7UUnit 45Units 30VVoltagesupply16, 71WW
78NotesSOLITRAC 31 • 4 … 20 mA/HART - four-wire40090-EN-131119
79NotesSOLITRAC 31 • 4 … 20 mA/HART - four-wire40090-EN-131119
83 Product descriptionSOLITRAC 31 • 4 … 20 mA/HART - four-wire40090-EN-131119• Modicationstatus,electronicsasof-01Theinstrumentisavailablein
Printing date:VEGA Grieshaber KGAm Hohenstein 11377761 SchiltachGermany40090-EN-131119All statements concerning scope of delivery, application, practi
93 Product descriptionSOLITRAC 31 • 4 … 20 mA/HART - four-wire40090-EN-131119The delivery must be checked for completeness and possible transit damage