Vega VEGABAR 65 4 … 20 mA_HART User Manual Page 55

  • Download
  • Add to my manuals
  • Print
  • Page
    / 76
  • Table of contents
  • BOOKMARKS
  • Rated. / 5. Based on customer reviews
Page view 54
HART output values
- HART value (Prima
ry Value) Process pressure
- HART value (Secondary Value) Temperature
Signal resolution 1.6 µA
Failure signal current output (adjustable) mA-value unchanged 20.5 mA, 22 mA, < 3.6 mA
Max. output current 22 mA
Load see load diagram under Power supply
Fullled NAMUR recommendations NE 43
Dynamic behaviour output
Run-up time approx. 10 s
90 %
100 %
10 %
t
t
T
A
t
t
S
2
1
Fig. 29: Sudden change of
the process variable, dead time t
T
, rise time t
A
and step response time t
S
1 Process variable
2 Output signal
Dead time 150 ms
Rise time 100 ms (10 90 %)
Step response time 250 ms (ti: 0 s, 10 90 %)
Damping (63 % of the input variable) 0 999 s, adjustable
Input variable
Adjustment
Adjustment range of the min./max. adjustment relating to the nominal measuring range:
- percentage value -10 110 %
- pressure value -20 120 %
VEGABAR 65 4 20 mA/HART 55
10 Supplement
27531-EN-081209
Page view 54
1 2 ... 50 51 52 53 54 55 56 57 58 59 60 ... 75 76