HART output values
- HART value (Prima
ry Value) Process pressure
- HART value (Secondary Value) Temperature
Signal resolution 1.6 µA
Failure signal current output (adjustable) mA-value unchanged 20.5 mA, 22 mA, < 3.6 mA
Max. output current 22 mA
Load see load diagram under Power supply
Fulfilled NAMUR recommendations NE 43
Dynamic behaviour output
Run-up time approx. 10 s
Fig. 29: Sudden change of
the process variable, dead time t
T
, rise time t
A
and step response time t
S
1 Process variable
2 Output signal
Dead time ≤ 150 ms
Rise time ≤ 100 ms (10 … 90 %)
Step response time ≤ 250 ms (ti: 0 s, 10 … 90 %)
Damping (63 % of the input variable) 0 … 999 s, adjustable
Input variable
Adjustment
Adjustment range of the min./max. adjustment relating to the nominal measuring range:
- percentage value -10 … 110 %
- pressure value -20 … 120 %
VEGABAR 65 • 4 … 20 mA/HART 55
10 Supplement
27531-EN-081209